First developed in the 1960s, X-ray photoelectron spectroscopy (XPS) has become a standard method in materials science. But now researchers at Linköping University in Sweden have shown that the method is often used erroneously.

"It is, of course, an ideal in research that the methods used are critically examined, but it seems that a couple of generations of researchers have failed to take seriously early signals that the calibration method was deficient," says Grzegorz Greczynski, senior lecturer in the Department of Physics, Chemistry and Biology at Linköping University. "This was the case also for a long time in our own research group. Now, however, we hope that XPS will be used with greater care."

Grzegorz Greczynski and his colleague Lars Hultman have shown that XPS can give misleading analysis results due to an erroneous assumption during calibration. They report their results in a paper in Scientific Reports.

XPS is used to determine the chemical composition of materials by irradiating them with X-rays and then taking advantage of the photoelectric effect, whereby the X-rays excite the electrons in the material so they are emitted as photoelectrons. The number of photoelectrons of a given energy that are detected by a measuring instrument can reveal the identity of the elements present in the material, as well as determine the chemical bonding between atoms and their proportional abundances.

More than 12,000 scientific articles with results obtained by XPS are now published each year. The technique was developed during the 1960s by Kai Siegbahn at Uppsala University in Sweden to become a useful and powerful method for chemical analysis, and the work led to him being awarded the Nobel Prize in physics in 1981.

"The pioneering work that led to the Nobel Prize is not in question here," says Hultman. "When the technique was developed initially, the error was comparatively small, due to the low calibration precision of the spectrometers used at the time. However, as spectroscopy developed rapidly and spread to other scientific fields, the instruments have been improved to such a degree that the underlying error has grown into a significant obstacle to future development."

What the researchers have discovered is that the original method is being used wrongly, through an erroneous assumption used in the calibration process. When calibrating the experiment, the signal from elemental carbon accumulating on the sample surface is often used.

It turns out that the carbon-based compounds that form naturally by condensation onto most samples give rise to signals that depend on the surroundings and the substrate onto which they are attached – in other words, the sample itself. This means that a unique signal is not produced, and large errors arise when a more or less arbitrary value is used as reference to calibrate the measuring instrument.

Criticisms about the method were raised as early as the 1970s and the 1980s. After that, however, knowledge about the error fell into obscurity. Hultman suggests that several factors have contributed to the error passing unnoticed for nearly 40 years. He believes that the dramatic increase in the number of journals that digital publishing has made possible is one such factor, while deficient reviewing procedures are another.

"Not only has a rapidly growing number of scientists failed to be critical, it seems that there is a form of carelessness among editors and reviewers for the scientific journals," he says. "This has led to the publishing of interpretations of data that are clearly in conflict with basic physics. You could call it a perfect storm. It’s likely that the same type of problem with deficient critical assessment of methods is present in several scientific disciplines, and in the long term this risks damaging research credibility."

Greczynski hopes that their discovery can not only continue to improve the XPS technique, but also contribute to a more critical approach within the research community.

"Our experiments show that the most popular calibration method leads to nonsense results and the use of it should therefore be terminated," Greczynski says. "There exist alternatives that can give more reliable calibration, but these are more demanding of the user and some need refinement. However, it is just as important to encourage people to be critical of established methods in lab discussions, in the development department, and in seminars."

This story is adapted from material from Linköping University, with editorial changes made by Materials Today. The views expressed in this article do not necessarily represent those of Elsevier. Link to original source.